Create mode– the default mode when you create a requisition and PunchOut to Bio-Rad. You can create and edit multiple shopping carts
Edit mode– allows you to edit or modify an existing requisition (prior to submitting). You will be able to modify only the cart that you have PunchedOut to, and will not have access to any other carts
Inspect mode– when you PunchOut to Bio-Rad from a previously created requisition but without initiating an Edit session, you will be in this mode. You cannot modify any Cart contents
CFX Opus Dx Real-Time PCR Detection Systems for In Vitro Diagnostics (IVD)
CFX Opus Dx Real-Time PCR Systems combine the flexibility you need in your workflow and outstanding performance to deliver reliable results that your clinical laboratory demands.
CFX Opus 96 Dx, CFX Opus 384 Dx,and CFX Opus Deepwell Dx are open systems offering the ultimate flexibility in commercial assay selection or rapid assay development in a clinical laboratory setting. CFX Opus 96 Dx and CFX Opus Deepwell can multiplex up to 5 targets per well and CFX Opus 384 Dx can multiplex up to 4 targets per well. CFX Opus Dx portfolio provides options to address multiple thruputs and sample volumes. The instruments offer precise quantification, accurate thermal cycler performance and powerful easy-to-use software.
Multiplex up to 5 targets per sample well on the CFX Opus 96 Dx and CFX Opus Deepwell Dx. The CFX Opus 384 can detect up to 4 targets per sample well. CFX Opus 384 Dx provides our highest throughput while CFX Opus Deepwell Dx accommodates larger sample volumes in a 96-well format. Optimize assay development with the thermal gradient function. Analyze and generate traceable, audit-ready reports with electronic signature functionality.
Licensed for PCR
Yes
Dimensions (WxDxH)
33 x 56 x 36 cm / 13 x 22 x 14 in
Weight
22 kg / 48 lb
Touch screen
Adjustable viewing angle from 15° to 22°C
Lid temperature range
Lid temperature range 30–110°C
Sample block type
Fixed 96 well
Fixed 384 well
Fixed 96 well
Consumables
96-well plates Single tubes, strip tubes
384-well plates
96-well plates Single tubes, strip tubes
Reaction volumes
1–50 µL (10–50 µL recommended)
1–30 µL (5–30 µL recommended)
10–125 µL
Maximum ramp rate
5°C/sec
2.5°C/sec
2.5°C/sec
Average ramp rate
3.3°C/sec
2°C/sec
2°C/sec
Thermal accuracy
±0.2ºC
±0.2ºC
±0.2ºC
Thermal uniformity
±0.3ºC
±0.3ºC
±0.4ºC
Gradient
Operational temperature range: 30–100°C Programmable span: 1–24°C
Sensitivity
1 copy of target sequence in human genomic DNA
Dynamic range
10 orders of magnitude
Excitation / Emission Ranges
450–684 nm / 510–730 nm
450–650 nm / 515–690 nm
450–684 nm / 510–730 nm
Calibrated fluorophores
Cal Gold 540, Cal Orange 560, Cal Red 610, Cy 5, Cy 5.5, FAM, HEX, Quasar 670, Quasar 705, ROX, Texas 615, VIC, SYBR, Texas Red
Cal Gold 540, Cal Orange 560, Cal Red 610, Cy 5, FAM, HEX, Quasar 670, ROX, Texas 615, VIC, SYBR, Texas Red
Cal Gold 540, Cal Orange 560, Cal Red 610, Cy 5, Cy 5.5, FAM, HEX, Quasar 670, Quasar 705, ROX, Texas 615, VIC, SYBR, Texas Red
Scan time
All channels: 12 sec FAM/SYBR: 3 sec
All channels: <20 sec FAM/SYBR: 8 sec
All channels: 12 sec FAM/SYBR: 3 sec
Multiplex
Up to 5 targets per well
Up to 4 targets per well
Up to 5 targets per well
Operating system
Windows 10 (64-bit)
Instrument control
Connect up to 4 CFX Opus Dx systems (mix of CFX Opus 96 Dx and CFX Opus 384 Dx) directly via USB cable or through laboratory network via Ethernet cable
Laboratory Informahion System compatible
Yes
Automation ready
API
21CFR Part 11
Assists with 21CFR Part 11 compliance through integrated secure username and password log-in, traceable audit-ready reports with electronic signature functionality
Data analysis
Allelic discrimination, End point, Gene expression, Quantification
CFX Opus 96 DX System, CFX Opus 384 Dx System,or CFX Opus Deepwell Dx System with CFX Maestro Dx SE Software are intended to perform fluorescence-based PCR to detect and quantitate nucleic acid sequences. The systems and software are intended for in vitro diagnostic use by trained laboratory technicians. The systems are intended to be used with third-party diagnostic nucleic acid tests which have been manufactured and labeled for diagnostic purposes.
USA:For In Vitro Diagnostic Use. The CFX Opus 96 Dx, CFX Opus 384 Dx, and CFX Opus Deepwell Dx Systems meet the requirement for distribution as an in vitro diagnostic device under the US FDA.
Canada:For In Vitro Diagnostic Use. The CFX Opus 96 Dx, CFX Opus 384 Dx, and CFX Opus Deepwell Dx Systems meet the requirement for distribution as an in vitro diagnostic device under Health Canada.
Europe:For In Vitro Diagnostic Use. The CFX Opus 96 Dx, CFX Opus 384 Dx, and CFX Opus Deepwell Dx Systems meet the requirement of the European In Vitro Diagnostic Regulation (2017/746).
Singapore:For In Vitro Diagnostic Use. The CFX Opus 96 Dx, CFX Opus 384 Dx, and CFX Opus Deepwell Dx Systems meet the requirements for distribution as an in vitro diagnostic instrument.
関连资料
Login Required
Number描述Options
7456
CFX Opus 96 Dx Real-Time PCR System Product Information Sheet