Sterilization filter of hospital vacuum system
【5】 Our products are original replacement parts, and other brands involved by our company have no brand significance, but are used as a reference for product models and customer selection.
More and more new coronavirus infection routes have been found in Wuhan. Hospitals should pay attention to whether there are sterilization facilities at the exhaust port of the negative pressure suction system.
Hangzhou Jiajie Electromechanical Equipment Co., Ltd. provides customers with customized vacuum solutions, which can meet the required performance at low cost based on full consideration of vacuum technical requirements. The hospital's negative pressure suction unit (system) exhaust disinfection and sterilization device, medical vacuum exhaust disinfection and sterilization device, special vacuum pump exhaust treatment for respiratory system, additional disinfection facilities at the exhaust outlet, Hangzhou Jiajie negative pressure suction sterilization filter passed the third-party 304 stainless steel food grade shell material and sterilization efficiency testing certification, The discharged gas complies with GB50751-2012 Technical Code for Medical Gas Engineering.
The installation template is as follows:
![](http://img3.59137.com/big/jushangyi/2020/11/24/10/0861136978893139.jpg)
1、 Check whether the negative pressure suction system is damaged due to leakage.
2、 Check whether the original sterilization filter used in the exhaust facility of the vacuum suction system used by the hospital has sterilization and disinfection effect on Wuhan COVID-19.
3、 If the vacuum system used is installed as required, exhaust disinfection treatment facilities shall be added for exhaust.
4、 Sterilization facilities require that all germs be killed instantly, and all germs be killed before being discharged to the atmospheric environment.
5、 The infectivity of the virus has exceeded the original standard.
Do not let the exhaust port of the negative pressure suction system become a pollution source!!!
Remarks: GB50751-2012 Technical Code for Medical Gas Engineering
Article 4.4.4
1. When several vacuum pumps share exhaust pipes, isolation measures shall be taken for exhaust of each vacuum pump;
2. The exhaust pipe shall be made of corrosion-resistant materials, and protective measures shall be taken. The lower part of the exhaust pipe shall be equipped with a drain valve;
3. The exhaust of vacuum pump should meet the requirements of hospital environmental health standards, and the exhaust port should be set with warning signs for harmful gases;
4. The exhaust outlet should be located outdoors, not at the same height as the medical air inlet doors and windows , other The distance between openings shall not be less than 3m;
5. The gas emission at the exhaust outlet shall not be affected by the monsoon, nearby buildings, terrain and other factors, and the discharged gas shall not be transferred to the working or living areas of other personnel.