FTC320EX hydrogen in oxygen analyzer

You added 1 item View Cart
Brief introduction
The FTC320 hydrogen in oxygen analyzer of Messkonzept GmbH in Germany | The oxygen in hydrogen analyzer adopts the principle of thermal conductivity. The measuring range of binary gas can reach 0-100% Vol, the resolution is 1ppm, the linearity can reach 1% FS, and the response time is very fast, up to 1 second (T90). FTC320 can be used for binary gas in a specific range. The thermal conductivity chip inside the sensor can be measured in high humidity and corrosive gas after special film processing.

FTC320EX hydrogen in oxygen analyzer

Detailed description of

FTC320EX Hydrogen in oxygen analyzer | Oxygen analyzer in hydrogen



Hydrogen production by electrolysis of water refers to the dissociation of water molecules under the action of direct current to generate oxygen and hydrogen, which separate from the anode and cathode of the electrolytic cell. According to the different membrane materials of electrolyzer, it can be divided into three technical routes: alkaline electrolysis (AWE), proton exchange membrane (PEM) electrolysis and solid oxide (SOEC) electrolysis.
Proton exchange membrane (PEM) electrolysis of water for hydrogen production: PEM electrolysis of water for hydrogen production selects proton exchange membrane with good chemical stability, proton conductivity and gas separation as solid electrolyte, which can effectively prevent electron transfer and improve the safety of the electrolytic cell.
Principle of PEM electrolyzing water to produce hydrogen: the main components of PEM electrolyzer are proton exchange membrane, anode and cathode catalytic layer, anode and cathode gas diffusion layer, anode and cathode terminal plate, etc. Among them, the diffusion layer, catalytic layer and proton exchange membrane constitute the membrane electrode, which is the main site for the material transport and electrochemical reaction of the whole water electrolysis cell. The characteristics and structure of the membrane electrode directly affect the performance and life of the PEM water electrolysis cell.

Germany Messkonzept GmbH company Of FTC320EX Hydrogen in oxygen analyzer | Oxygen analyzer in hydrogen The measuring range of binary component gas can reach 0-100%Vol , resolution is 1ppm, Linear reachability 1%FS , very fast response time one Seconds( T90 )。 FTC320 Can be in a specific range of binary gas. Sensor internal thermal conductivity chip After special film treatment, it can be measured in high humidity and corrosive gas.


Remarks: FTC320EX It is the version with explosion-proof mark

FTC320 hydrogen in oxygen analyzer | oxygen in hydrogen analyzer characteristic:

l   six Power linear fitting, very high precision.

l   Isolated 4~20mA Output, in 100%Vol. The range can be set arbitrarily within the range of

l   Free setting start and end

l   classical two Point calibration or single point calibration

l   Display screen can be set freely ppm or Vol.% Display, resolution up to 1ppm 

l     Rapid response, T90 Response time is less than one second ( Depends on flow )

l   Positive pressure resistance (20bar) And vacuum sealed stainless steel (LF316i)  Gas pipeline

l   The three isolation relays display alarm and instrument status respectively

l   RS232 Digital interface, which can output all parameters

l   Throughout 100%Vol. In scope Digital output reaches 1ppm resolving power

l     Small and sturdy transmitter embedded in aluminum housing, suitable for field operation (protection class: IP65

l     Size: 145mm( wide ) x 80mm ( High, no Connector )x 85 mm( deep )

l   Power Supply: 18V-36VDC/700mA


Explosion proof glass bead filling option

Flammable and explosive gases may be introduced into the FTC320 In the analyzer. Here, the inner space of the shell is densely covered with glass beads (diameter 0.6mm )。 In the case of explosive environment caused by unlikely leakage, the small space between glass beads prevents simultaneous ignition caused by analyzer failure.

Do not open the glass bead when it is loose FTC320 Housing. When the shell of the analyzer is opened, it may not work safely with flammable gas, causing explosion.

Anti condensation and dust handling options

When the desiccant or other drying equipment and particle filtration fail or have low efficiency, the measured gas may condense or large particles enter the analyzer, causing damage to the analyzer. The sensor inside the analyzer adopts one layer PTFE The coating protects the thermal conductivity sensor.


FTC320 hydrogen in oxygen analyzer | oxygen in hydrogen analyzer measuring range

Measuring gas

Background gas

Basic range

Min. Range

Min. Compression zero range

H2 , Hydrogen

O2 , Oxygen

0~100% Vol.

0~0.5% Vol.

98%~100%

O2 , Oxygen

H2 , Hydrogen

0~100% Vol.

0~3% Vol.

99.5%~100%

 

  FTC320 hydrogen in oxygen analyzer | oxygen in hydrogen analyzer technical parameter

Measuring gas

Basic range

linear

1%FS

Repeatability

< Range 1%

noise

< Min. Range 1%

Zero drift

< Min. Range 2%/ week

Error (ambient temperature change)

< Min. Range 1%/10

Error (flow change @80l/h

< Min. Range 1%/10l/h

Error (pressure change, higher than 80kPa

< Min. Range 1%/1kPa

Warm up time

about thirty minute; one Hour (small range)

working temperature

-5~+50

flow

40l/h~150l/h; 60l/h~80l/h( recommend )

Response time( T90

<1s @60l/h

Gas pressure (absolute pressure)

recommend 1bar , Max. withstand voltage 20bar

Size (without connector) / weight

145mm( wide ) x 80mm ( high )x 85 mm( deep )/1800g( Max. )

Supply voltage

18V-36VDC/700mA

Installation mode

Wall mounting

FTC320 hydrogen in oxygen analyzer | oxygen in hydrogen analyzer Overall dimension


FTC320 hydrogen in oxygen analyzer | oxygen in hydrogen analyzer Electrical connection and grounding


Connectors and cables:( seven hundred and twelve IP67 )With extruded connector plug, length: two M, the cable has an open end. Cable A and C The cross sectional area of the middle conductor is zero point one four Mm2, cable B by zero point two five square millimetre.

Warning!! Hydrogen is explosive gas, please do not disassemble or install the analyzer with electricity!!